Endurity Core pacemaker (Abbott)
Endurity Core is a pacemaker family from Abbott, formerly St. Jude Medical, with two models: the single-chamber PM1152 and the dual-chamber PM2152. Both weigh 19 g, use an IS-1 connector and allow 1.5 T MRI under conditions when implanted with the leads the manufacturer specifies. The price depends on the model, the leads and the quantity.
- PM1152 single-chamber
- PM2152 dual-chamber
- IS-1
- 1.5 T MRI with conditions
Updated: 20 November 2026
Models and specifications
| Model | Type | Size (H × W × T) | Weight | Volume | Pacing modes |
|---|---|---|---|---|---|
| PM1152 | Single-chamber (SR) | 41 × 50 × 6 mm | 19 g | 9.7 cc | VVI(R), VVT(R), VOO(R), AAI(R), AAT(R), AOO(R) |
| PM2152 | Dual-chamber (DR) | 46 × 50 × 6 mm | 19 g | 10.41 cc | DDD(R), DDI(R), DVI(R), DOO(R), VDD(R) and single-chamber modes |
Both models use inductive telemetry: they are interrogated and programmed by placing the head of Abbott's Merlin programmer over the pacemaker. The hospital needs that programmer for the implant and follow-up.
Main features
- AutoCapture: confirms beat by beat that each ventricular pulse captures and, if it does not, delivers a 5.0 V backup pulse. This lets the output stay low, which extends the battery.
- VIP (Ventricular Intrinsic Preference), on the dual-chamber PM2152: lets the patient's own conduction work to avoid unnecessary right-ventricular pacing.
- AF Suppression algorithm, for patients with an atrial lead.
- Stored electrograms of up to 2 minutes and, on the PM2152, programmable atrial tachyarrhythmia (AT/AF) alerts.
- Lead monitoring with automatic polarity switching.
MRI
According to its spec sheets, Endurity Core allows 1.5 T MRI under conditions when the whole system uses Tendril STS 2088TC leads (46, 52 or 58 cm) or IsoFlex Optim 1944 or 1948. The conditions include a whole-body SAR of up to 2 W/kg and an exclusion zone: the scan isocentre must be below L4 or above C1, and with the 2088TC at least 10 cm above C1. Abbott offers a handheld activator that switches the pacemaker to MRI mode before the scan and back afterwards.
If the patient may need a full-body or 3 T MRI, Abbott has models such as Endurity MRI and Assurity MRI; we describe them in Abbott pacemakers. The current conditions for each combination are in Abbott's MRI-Ready Systems manual.
Compatible leads
| Lead | Fixation | Minimum introducer | Lengths |
|---|---|---|---|
| Tendril STS 2088TC | Active (extendable and retractable helix) | 6 F | 46, 52 and 58 cm (MR Conditional); 65 and 100 cm |
| IsoFlex Optim 1944 (J-shaped) | Passive (tines) | 7 F | 46 y 52 cm |
| IsoFlex Optim 1948 (straight) | Passive (tines) | 7 F | 52 y 58 cm |
All are bipolar with an IS-1 connector. A single-chamber system uses one lead and a dual-chamber system uses two, one in the atrium and one in the ventricle. More about the system in generator and leads.
Battery life
The PM1152 spec sheet states 14.4 years of service life with 100% VVI pacing at 60 bpm, 2.5 V, 0.4 ms and 500 Ω, with AutoCapture off. There are 6 months between the elective replacement indicator (ERI) and end of life (EOL). On the PM2152 the battery life depends on how both chambers are programmed; check the current spec sheet. We explain what it depends on in how long a pacemaker lasts.
Endurity Core price
The price depends on:
- The model: single-chamber PM1152 or dual-chamber PM2152.
- The leads: one or two, Tendril STS or IsoFlex Optim, and their length.
- The introducers and implant accessories.
- The quantity: the unit price falls with volume.
Volume pricing
The unit price falls as the quantity grows. That is why we do not publish a single price: we calculate it from the model, the quantity and how often you buy.
Unit price (illustrative)
The pacemaker type, brand and programming are decided by the cardiologist or electrophysiologist according to each patient's diagnosis. This page informs about options and prices and does not replace a medical assessment.
Frequently asked questions
What is the price of the Endurity Core pacemaker?
It depends on whether it is the single-chamber PM1152 or the dual-chamber PM2152, the leads that go with it and the quantity. Send us the model and number of units for a quote.
Is Endurity Core MRI compatible?
Yes, under conditions. According to its spec sheets, it allows 1.5 T MRI with Tendril STS 2088TC or IsoFlex Optim 1944 or 1948 leads and with an exclusion zone for the isocentre. Abbott's current MRI-Ready Systems manual must be checked before each scan.
What is the difference between Endurity Core and Endurity MRI?
They are the same size. Endurity MRI is designed for full-body 1.5 T or 3 T MRI with Abbott MRI leads. Endurity Core, according to its spec sheets, allows 1.5 T MRI with an exclusion zone.
Which lead is used with Endurity Core?
The spec sheets list the Tendril STS 2088TC, with active fixation, and the IsoFlex Optim 1944 and 1948, with passive fixation. All are bipolar with an IS-1 connector.
Is Endurity Core a St. Jude pacemaker?
Yes. It was launched under the St. Jude Medical brand, which Abbott bought in 2017. Today it is sold as Abbott.
Request a quote
Tell us the brand, type and quantity. We reply within 24 hours with the unit price based on volume, availability and delivery time.
- Unit price based on the total quantity.